| Home | E-Submission | Sitemap | Contact Us |  
top_img
J Korean Soc Ther Radiol Oncol > Volume 16(2); 1998 > Article
The Journal of the Korean Society for Therapeutic Radiology and Oncology 1998;16(2): 195-202.
Measurement and Monte Carlo Simulation of 6 MV X-rays for Small Radiation Fields
Dong Hyeok Jeong, Jeong Ok Lee, Jeong Ku Kang, Soo Kon Kim, Seung Kon Kim, Sun Rock Moon
1Department of Radiation Oncology, Wonkwang University School of Medicine, Korea.
2Department of Radiotechnology, Wonkwang Health Science College, Korea.
3Department of Radiation Oncology, Presbyterian Medical Center, Korea.
4Department of Physics, Chonbuk National University, Korea.
ABSTRACT
PURPOSE:
In order to obtain basic data for treatment plan in radiosurgery, we measured small fields of 6 MV X-rays and compared the measured data with our Monte Carlo simulations for the small fields.
MATERIALS AND METHODS:
The small fields of 1.0, 2.0 and 3.0 cm in diameter were used in this study. Percentage depth dose (PDD) and beam profiles of those fields were measured and calculated. A small semiconductor detector, water phantoms, and a remote control system were used for the measurement. Monte Carlo simulations were performed using the EGS4 code with the input data prepared for the energy distribution of 6 MV X-rays, beam divergence, circular fields and the geometry of the water phantoms.
RESULTS:
In the case of PDD values, the calculated values were lower than the measured values for all fields and depths, with the differences being 0.3 to 5.7% at the depths of 2.0 to 20.0 cm and 0.0 to 8.9% at the surface regions. As a result of the analysis of beam profiles for all field sizes at a depth of 10cm in water phantom, the measured 90% dose widths were in good agreement with the calculated values, however, the calculated penumbra radii were 0.1 cm shorter than measured values.
CONCLUSION:
The measured PDDs and beam profiles agreement with the Monte Carlo calculations approximately. However, it is different when it comes to calculations in the area of phantom surface and penumbra because the Monte Carlo calculations were performed under the simplified geometries. Therefore, we have to study how to include the actual geometries and more precise data for the field area in Monte Carlo calculations. The Monte Carlo calculations will be used as a useful tool for the very complicated conditions in measurement and verification.
Key Words: Small field, Semiconductor detector, Monte Carlo simulation, EGS4
TOOLS
PDF Links  PDF Links
Full text via DOI  Full text via DOI
Download Citation  Download Citation
Share:      
METRICS
1,577
View
30
Download
Related articles
Understanding the Response Characteristics of X-ray Verification Film1998 December;16(4)
Transmission Dose Estimation Algorithm for Irregularly Shaped Radiation Field2002 September;20(3)
Editorial Office
Department of Radiation Oncology, Samsung Medical Center,
Proton Therapy Center, B2, 81, Irwon-ro, Gangnam-gu, Seoul 06351, Republic of Korea
Tel : +82-2-3410-3617
E-mail: rojeditor@gmail.com, roj@kosro.or.kr
Copyright © The Korean Society for Radiation Oncology.                      Developed in M2PI
Close layer
prev next